When an EKG Passes Inspection but Still Doesn't Work: What PBES's 2026 Service Records Reveal
A descriptive look at 291 EKG inspection records and 30 EKG service items from PBES field work — and why a clean simulator test doesn't always mean the patient connection works.

Most EKG problems that reach a service technician start with one sentence from clinical staff: “It keeps saying leads off,” or “the tracing is noisy.” A technician connects a patient simulator, the machine prints a clean waveform, and the inspection is recorded as Pass. Both things can be true at the same time.
This report looks at what PBES's own 2026 field records show about that gap. It is a descriptive review of recorded service and inspection work from June 2 through November 18, 2026 — year-to-date findings, not a completed annual report, not a clinical study, and not a judgment on any manufacturer's equipment.
Executive findings
- EKG service items
- 30
- Across 26 service sessions, June 2 – Nov 18
- EKG inspection records
- 291
- Logged in the same window
- Unique EKG devices
- 254
- Some devices were inspected more than once
- Pass records with a failed accessory
- 2
- Two records for one device, on two visits — not two faulty devices
- **291 EKG inspection records** across **254 unique devices** were logged in the window: 239 Pass, 45 retired during the session, 4 Fail, 2 excluded and 1 pending.
- Of the 239 Pass records, **145 explicitly record a functional pass** and **104 record a simulator heart rate (BPM)**. The other Pass records leave those fields blank, so the data cannot show whether those checks were performed or simply not entered.
- **Two Pass records for one device** — on two separate visits — recorded a failed accessory: a broken trunk-cable-to-lead clip. The overall status and the accessory finding disagreed. This is a documentation discrepancy, not a simulator failure.
- Across **30 EKG service items in 26 service sessions**, the recurring themes were cables and lead wires, the electrode interface, battery and power, the print workflow, and settings or lead-placement warnings.
Why static tests may miss patient interface problems
An EKG patient simulator feeds the machine a known, stable signal through the lead set. It is the right tool to confirm that the EKG can acquire, process and print a waveform at the expected rate. It is a different instrument from an electrical safety analyzer, which checks leakage current and ground integrity. A device can pass one and still need attention from the other.
What a still, bench-top simulator connection does not fully reproduce is the patient side: adhesive electrodes on skin, clips that have been stored clamped to electrodes, retention clips that no longer hold, a trunk cable that only fails when it bends, and a patient who moves. A connection that is perfect on the bench can behave differently on a person.
Patient simulator, held still
- Simulator
- Lead set
- EKG
Real electrodes, real movement
- Electrode (failure point seen in 2026 records)
- Clip (failure point seen in 2026 records)
- Lead wire (failure point seen in 2026 records)
- Trunk cable (failure point seen in 2026 records)
- EKG
That is not an argument against simulator testing. In one November visit, a technician isolated a cut, open trunk cable on a different ELI 280 by swapping in a known-good cable assembly — simulator-based troubleshooting found the fault cleanly. The point is that a single static pass is one layer of evidence, not the whole picture.
Three anonymized field cases
Facility names, locations, serial numbers and control numbers are left out. Each case states what the service record documents, and no more.
Case A — “Leads off” that would not reproduce
In early November, staff reported a recurring Leads Off message on a Burdick ELI 280. The technician could not reproduce it: with the simulator connected, the unit printed a clean waveform. Two observations went into the record — the clips had been stored attached to electrodes, and the battery was depleted. An adhesion or electrode-interface problem was suspected but not proven.
Case B — Tested correctly, then artifact on a patient
Across two November visits, another ELI 280 tested correctly but staff reported artifact once it was connected to a patient. Follow-up found worn lead wires, a broken retention clip and movement-related noise, and the technician moved the unit to a known-good lead set. The records document those findings; on their own they do not prove the problem is permanently resolved. A related PBES write-up: Burdick ELI 280 lead reading artifact repair.
Case C — One lead, one flex point
In August, a Welch Allyn CP150 showed an intermittent V3 lead. Flexing the trunk cable changed the V3 signal, the cable fault was confirmed and replacement was recommended. No static simulator pass was explicitly documented in that record. Read the related case: Welch Allyn CP150 V3 lead fault and trunk cable repair.
- 1Electrode & adhesionClips stored attached to electrodes; adhesion suspected, not proven (Case A).
- 2Clip / snapBroken trunk-cable-to-lead clip recorded on two Pass inspections of one device.
- 3Lead wiresWorn lead wires with movement-related noise (Case B).
- 4Retention clipBroken retention clip found after artifact on a patient (Case B).
- 5Trunk cableFlex-dependent V3 fault (Case C); a cut, open trunk isolated with a known-good assembly.
- 6Device & powerDepleted battery noted alongside a Leads Off complaint (Case A).
Inspection records show the same tension from the other direction. One October inspection noted a strong simulator signal with no artifacts — alongside staff complaints about poor connections and a V1/V2 reversal warning — and was recorded Pass. Another, on a GE MAC 2000, was recorded Pass with a 60 BPM simulator entry and a documented printout alignment and paper issue.
Five key categories in the records
Grouping the 2026 EKG findings by where the problem sat, rather than by model, produces five practical categories. They are themes, not a frequency ranking.
- 01Cables, lead wires & intermittent connections
Trunk cables that fail on flex, worn leads, broken clips and retention clips.
- 02Electrode & patient interface
Adhesion, electrode storage habits and motion artifact once a patient is connected.
- 03Battery & power
Depleted batteries noted alongside connection complaints.
- 04Print workflow
Paper loading and printout alignment issues recorded on otherwise passing units.
- 05Settings, interpretation & lead placement
Configuration questions and warnings such as a V1/V2 reversal alert.
Comprehensive verification vs. a single pass/fail
An overall status is a summary. When the summary says Pass and the accessory field says Fail — as it did for one device on two visits this year — the facility and the next technician both need to see the detail, not just the headline.
- 239Pass
- 45Retired during session
- 4Fail
- 2Excluded
- 1Pending
| Status | Records |
|---|---|
| Pass | 239 |
| Retired during session | 45 |
| Fail | 4 |
| Excluded | 2 |
| Pending | 1 |
| Total | 291 |
| Pass records with functional pass recorded | 145 of 239 |
| Pass records with simulator BPM recorded | 104 of 239 |
The second chart matters most. 104 simulator heart-rate entries among 239 passes does not mean the rest were never tested; it means the record cannot show it either way. A blank field is not evidence of a completed test. That is why PBES is moving EKG work toward a layered verification, with each layer recorded on its own.
- Layer 1Electrical safety
Safety analyzer checks leakage and ground — a separate instrument from the EKG simulator.
- Layer 2Function on the simulator
Known signal in, clean waveform and correct rate out. The BPM is written down, not assumed.
- Layer 3Accessories under flex
Lead wires, clips, retention clips and trunk cable inspected and flexed while watching the trace.
- Layer 4Patient interface & workflow
Electrode storage, battery, paper and print alignment, settings and lead-placement warnings — checked against what staff reported.
- Layer 5A record that shows each layer
Every result in its own field, so an accessory Fail stays visible even when the overall status is Pass.
How operational data compounds
Every finding in this report exists because a technician wrote it down in TESSA, PBES's equipment and service platform. TESSA's guided troubleshooting is designed to help technicians capture what was reported, what was tested and what was found in a consistent structure, and to generate the service report from those entries.
The opportunity — and today it is an opportunity, not a measured result — is that patterns like passes on the simulator, fails on the patient become visible across devices and visits, and one technician's experience becomes available to the next. Over time that could make EKG service more consistent and shorten diagnosis. We will report that when we can measure it. The longer story is in Building TESSA.
Every Service Makes the Next One Smarter.
What facility managers can do
- Report symptoms the way staff see them — “leads off,” “noisy on V1,” “only when the patient moves” — and note when they happen.
- Store lead wires unclipped from electrodes, and don't leave electrodes attached between patients.
- Glance at clips, retention clips and the trunk cable during routine use; replace lead sets that are cracked, stiff or no longer hold.
- Keep a known-good spare lead set so staff can tell whether a problem follows the cable.
- Keep batteries charged and paper loaded and aligned — some “machine” complaints are power or print workflow.
- Ask your service provider for accessory, functional and simulator results as separate lines on the inspection record, not only an overall status.
None of this replaces clinical judgment. If a tracing looks wrong, clinical staff should follow their facility's protocol. For hands-on help, see PBES EKG and ECG machine repair.
Methodology and limitations
- Source: a read-only review of PBES TESSA service and inspection records. No records were changed for this report.
- Window: June 2 – November 18, 2026. These are year-to-date findings, not a completed 2026 annual report.
- Scope: EKG/ECG devices only — 291 inspection records (254 unique device IDs) and 30 service items across 26 sessions.
- Design: descriptive analysis of recorded operations. Not randomized, not controlled, no comparison group.
- Blank fields are treated as unknown — neither as a test performed nor as a test skipped.
- No failure, false-negative or downtime rates are calculated, and no external industry statistics are cited.
- Cases are anonymized; models are named for technical context only. No patient information was used.
- Simulator results and electrical safety analyzer results are separate measurements and are reported separately.
Check the whole chain — machine, cables and connection.
PBES services EKG and ECG machines on site across Florida, including lead-set and trunk-cable inspection, simulator verification and electrical safety testing, with each result documented in TESSA.