PBES case studies
Real service write-ups from PBES technicians: reported issue, investigation, root cause, resolution, and verification for clinical equipment.
McKesson ELI280 Lead Off Error Repair Case Study
McKesson ELI280 ECG/EKG Machine
A McKesson ELI280 ECG machine reported a lead off error. Learn how PBES biomedical technicians evaluated the trunk cable connections to resolve the issue.
Civco MT-VL-1 Vacuum Compressor Air Supply Failure Repair
Civco MT-VL-1 Vacuum Compressor
PBES repair case study on a Civco MT-VL-1 vacuum compressor producing vacuum but failing to supply compressed air due to a burned contactor and relay.
MTI 527-115 Podiatry Chair Actuator Failure Repair
MTI 527-115 Podiatry chair
Diagnostic case study of an MTI 527-115 podiatry chair with failing lift, tilt, and back actuators and a missing lockout switch at a Florida surgery center.
MTI 527-115 Podiatry Chair Lift Actuator Repair Case Study
MTI 527-115 Podiatry Chair
PBES service case study on repairing an MTI 527-115 podiatry chair in a Florida hospital, replacing the lift actuator, friction brake, and hardware.
Tuttnauer 2340EA Autoclave Leak and Cycle Failure Repair
Tuttnauer 2340EA Autoclave
Case study on troubleshooting a Tuttnauer 2340EA autoclave experiencing front-door steam leaks, cycle aborts, and over-temperature errors in a hospital.
MTI 527-115 Podiatry Chair Actuator Diagnostic and Repair
MTI 527-115 Podiatry chair
Onsite biomedical diagnostic of an MTI 527-115 podiatry chair experiencing lift, tilt, and back movement failures attributed to failing actuators.
MTI 527-115 Podiatry Chair Actuator Failure Repair Case Study
MTI 527-115 Podiatry chair
Onsite biomedical diagnosis of an MTI 527-115 podiatry chair with failing lift, tilt, and back actuators, plus a missing electrical lockout switch.
GE OEC 9800 Plus C-Arm Foot Pedal Sticking Repair Case Study
GE OEC 9800 Plus C-Arm
PBES service report: Investigating intermittent foot pedal sticking and exposure overrun on a GE OEC 9800 Plus C-Arm at a Florida surgery center.
SciCan Statim 5000 G4 Sterilizer Maintenance Repair Case Study
SciCan Statim 5000 G4 Sterilizer
PBES technician completed scheduled preventive maintenance on a SciCan Statim 5000 G4 cassette autoclave at a Florida surgery center.
Tuttnauer 2340EA Autoclave Door Leak Repair Case Study
Tuttnauer 2340EA Autoclave
Case study on diagnosing and repairing front door leaks, cycle failures, and worn components on a Tuttnauer 2340EA autoclave in a Florida hospital.
MTI 527-115 Podiatry Chair Actuator and Hardware Repair
MTI 527-115 Podiatry Chair
PBES service report: Lift actuator, friction brake, and hardware repairs completed on an MTI 527-115 podiatry chair in a Florida hospital facility.
GE Venue R1 Ultrasound Power Failure Repair Case Study
GE Venue R1 Ultrasound Machine
PBES service case study: GE Venue R1 ultrasound failing to power on in a Florida hospital, diagnosed with a faulty internal RAM module.
Burdick ELI 280 Lead Reading Artifact Repair Case Study
Burdick ELI 280 ECG/EKG Machine
Case study on diagnosing lead reading issues and artifact on a Burdick ELI 280 ECG machine during a Florida biomedical service call.
Welch Allyn Spot Vital Signs LXi NIBP Reading Failure Repair
Welch Allyn LXI 43MTO Vital Signs Monitor
A clinic reported a Welch Allyn Spot Vital Signs LXi inflating without displaying blood pressure readings. PBES resolved the failure without replacement parts.
Olympus ESG-200 ESU Configuration and Foot Pedal Repair
Olympus ESG-200 Electrosurgical Unit
PBES technicians addressed configuration errors and diagnosed a faulty bipolar foot pedal on an Olympus ESG-200 electrosurgical unit in a surgery center.
Zoll R Series Defibrillator Battery Indicator Repair
Zoll R SERIES Defibrillator
PBES service report: Zoll R Series defibrillator battery indicator illuminated orange in a clinic hallway. Battery replaced and 30-joule discharge verified.
Midmark M9-022 Door Gasket Leak Repair Case Study
Midmark (Ritter) M9-022 Steam Sterilizer
PBES technician resolves front door steam leaks and unseated door gaskets on a Midmark M9-022 autoclave following unit relocation.
Welch Allyn Connex Spot Monitor Battery Repair Case Study
Welch Allyn Connex Spot Monitor Vital Signs Monitor
PBES technician evaluates a Welch Allyn Connex Spot Monitor failing to charge or hold power at a Florida urgent care center.
Tuttnauer TVET 10E Autoclave Maintenance Repair Case Study
Tuttnauer TVET 10E Autoclave - Steam Sterilizer
PBES completed scheduled preventive maintenance on a Tuttnauer TVET 10E steam sterilizer at a Florida hospital facility.
STERIS E3021-1 Autoclave Sight Glass Leak Repair Case Study
STERIS E3021-1 (Amsco Eagle 3023) Prevac Steam Sterilizer
PBES technicians resolved a hissing leak on a STERIS Amsco Eagle 3023 sterilizer at a Florida surgery center by isolating the steam generator sight glass assembly.
DEKA SmartXide2 Laser HVPS Coupling Repair Case Study
DEKA SMARTXIDE2 Laser system
PBES repaired a DEKA SmartXide2 laser system involving a damaged HVPS coupling, complete reassembly, functional testing, and return to service.
Steris Reliance 444 Washer Disinfector Maintenance & Repair
Steris (Amsco Reliance) 444 Washer/Disinfector Washer Disinfector
PBES service report detailing Inspection 1 quarterly preventive maintenance and chemical injection servicing on a Steris Reliance 444 Washer Disinfector.
MTI 527-115 Podiatry Chair Armrest Realignment Repair Case Study
MTI (Medical Technology Inc.) 527-115 Podiatry Chair
Read how PBES resolved a misaligned armrest on an MTI 527-115 podiatry chair in a hospital setting, restoring normal function without replacement parts.
SciCan STATIM 2000 Cycle Fault 25 Repair Case Study
SciCan STATIM 2000 Cassette Autoclave (Model No. 121101) Autoclave / Sterilizer
PBES technician resolves a SciCan STATIM 2000 Cycle Fault 25 caused by a failed water pump overheating the steam generator in a Florida hospital.
Welch Allyn 53NTO C7 Error Repair Case Study
Welch Allyn 53NTO vital-signs-monitor
PBES biomedical service report on troubleshooting an intermittent C7 error on a Welch Allyn 53NTO vital signs monitor in Florida.
Steris Amsco 400 Series Autoclave Door Jam Repair
Steris/Amsco AMSCO 400 Series (Model 20) Steam Sterilizer
PBES field service report: Resolving a door obstruction on a Steris Amsco 400 Series steam sterilizer at a Florida surgery center.
GE Brivo OEC 715 Relocation PM Repair Case Study
GE Hualun Medical Systems, Co. Ltd Brivo OEC 715 Mobile C-arm X-ray system
PBES completed post-relocation reassembly, preventive maintenance, and OEM spec testing on a GE Brivo OEC 715 mobile C-arm system in a Florida hospital.
Zoll R Series Broken Knob Repair Case Study
Zoll R SERIES Defibrillator
A PBES technician responds to a surgery center's report of a Zoll R Series defibrillator with a broken control knob. See the resolution in this repair case study.
Steris Amsco 5052 Washer Water Pressure Gauge Leak Repair
Steris/Amsco 5052 (REF FH14) Washer/Disinfector
A surgery center reported a water leak on a Steris Amsco 5052 washer/disinfector. PBES isolated the supply and replaced the leaking inlet water pressure gauge.
Cardinal Detecto 6550-C MedVue Scale Cable Repair
Cardinal Detecto 6550-C Medical Scale
PBES repaired broken serial cable connector pins and adjusted the loose display mount on a Cardinal Detecto 6550-C MedVue medical scale in Florida.
GE Brivo OEC 715 C-Arm PM Repair Case Study
GE Brivo OEC 715 Mobile C-Arm
Case study on preventive maintenance and FDA compliance inspection for a GE Brivo OEC 715 mobile C-arm at a Florida clinic.
Steris Amsco Eagle 3023 Sight Glass Leak Repair Case Study
Steris / Amsco Eagle 3023 Prevac Steam Sterilizer (nameplate E3021-1) Steam Sterilizer
PBES technicians resolved a leaking steam generator sight glass assembly on a Steris Amsco Eagle 3023 sterilizer at a Florida surgery center.
ZOLL AED Plus Post-Use Inspection and Testing Case Study
ZOLL MEDICAL, INC. AEDPLUS AED
PBES technician conducts post-use functional verification, pad replacement, and energy simulation on a ZOLL AED Plus at a Florida clinic.
Midmark Ritter M9 Burning Smell Repair Case Study
Midmark/Ritter M9-001 general
PBES repaired a Midmark Ritter M9 autoclave at a Florida hospital after staff reported a burning smell caused by a detached overtemp thermostat.
JPEX Medical M1040P EKG Failure Repair Case Study
JPEX MEDICAL M1040P Patient Monitor
A PBES technician resolves an EKG failure on a JPEX Medical M1040P patient monitor by correcting a lead configuration setting. Read the full repair report.
Tuttnauer EZ9 Autoclave Annual Service and Repair
Tuttnauer EZ9 Automatic Autoclave
Case study on annual service, chamber flushing, and plumbing system maintenance for a Tuttnauer EZ9 autoclave in a Florida clinic.
Medfusion 3500 Broken Plunger Lever Repair Case Study
Medfusion 3500 infusion-pump
Case study on repairing a Medfusion 3500 syringe infusion pump with a broken plunger lever at a Florida surgery center.
Stryker 1731 Renaissance Stretcher Hydraulic Leak Repair
Stryker Medical 1731 Stretcher
PBES service report details hydraulic pump inspection and replacement on a Stryker 1731 Renaissance stretcher at a Florida surgery center.
Tuttnauer EZ9 Autoclave Annual Service Maintenance Case Study
Tuttnauer EZ9 Automatic Autoclave
Case study of annual sterilizer preventive maintenance on a Tuttnauer EZ9 autoclave, including chamber cleaning and plumbing system flushing.
Tuttnauer 3870EA Autoclave System Flush Repair Case Study
Tuttnauer 3870EA Autoclave
Case study on repair services for a Tuttnauer 3870EA autoclave at a Florida surgery center, including chamber, reservoir, and plumbing flushes.
Tanita TBF-300 Scale Printer Issue Repair Case Study
Tanita TBF-300 Scale
PBES service report: Resolving faded and intermittent thermal printing on a Tanita TBF-300 body composition scale at a Florida clinic.
GE MAC 5500 ECG Issue Repair Case Study
GE MAC 5500 EKG
A Florida clinic reported ECG issues with pediatric patients on a GE MAC 5500. A PBES technician investigated and recommended replacing a worn trunk cable.
GE Mac 2000 Printer Jam Repair Case Study
GE Mac 2000 ECG / EKG Machine
A PBES technician diagnoses and resolves printer jams and misaligned printouts on a GE Mac 2000 ECG machine at a Florida surgery center. See the simple root cause.
Colin M30 Printer Error Repair Case Study
Colin M30 Patient Monitor
A PBES technician resolves a 'No Paper' error on a Colin M30 patient monitor by correcting improper paper loading. A real repair case study.
Zoll R Series Defibrillator Knob Repair Case Study
Zoll R SERIES Defibrillator
A PBES case study on a simple but critical repair of a faulty selector knob on a Zoll R Series defibrillator at a Florida clinic.
Welch Allyn 53NTO Battery & NIBP Repair Case Study
Welch Allyn 53NTO Patient Monitor
A real PBES repair case study on a Welch Allyn 53NTO patient monitor shutting down from a bad battery and reporting high NIBP readings.
Zoll R Series Defibrillator Battery Warning Repair Case Study
Zoll R SERIES Defibrillator
A PBES technician responds to a Zoll R Series defibrillator showing a battery warning. See the simple steps taken to resolve the issue and verify function.
Edan M3A NIBP Function Test: PM Case Study
EDAN INSTRUMENTS M3A Patient Monitor
A PBES technician investigates a reported blood pressure failure on an Edan M3A patient monitor. See the diagnostic and testing steps from the real service report.
Zoll R Series Defibrillator Battery Repair Case Study
Zoll R SERIES defibrillator
A PBES technician diagnoses and resolves a 'Battery on Orange' alert on a Zoll R Series defibrillator at a Florida surgery center. See the repair process.
Welch Allyn CP150 V3 Lead Fault Repair Case Study
Welch Allyn CP150 EKG
A PBES technician diagnoses an intermittent V3 lead fault on a Welch Allyn CP150 ECG. See how a failing trunk cable was identified as the root cause.
ZEISS 750i Lab Analyzer Startup Failure Repair Case Study
ZEISS 750I Lab Analyzer
A Florida clinic's ZEISS 750i Lab Analyzer failed to start up. See how a PBES tech diagnosed and resolved an incorrect BIOS date/time configuration without parts.
Zoll AED Plus Battery Replacement Repair Case Study
ZOLL AED PLUS Automated External Defibrillator
A PBES technician responds to a clinic's request for a battery replacement on a Zoll AED Plus. See the documented steps from this real service report.
Welch Allyn SPOT Vital Signs BP Issue Repair Case Study
Welch Allyn SPOT VITAL SIGNS Patient Monitor
A PBES technician diagnoses a reported BP issue on a Welch Allyn SPOT Vital Signs monitor, finding a damaged hose during inspection. See the diagnostic steps.
Welch Allyn SPOT Repair: SpO2 & NIBP Faults | PBES Case Study
Welch Allyn VITAL SIGNS SPOT Vital Signs Monitor
A Florida clinic's Welch Allyn SPOT monitor had constant SpO2 alarms and slow NIBP. See how a PBES technician diagnosed a damaged accessory.
Mortara Eli 250 Dim Screen & Charging Issue Repair Case Study
Mortara Eli 250 EKG
Read a PBES repair case study for a Mortara Eli 250 EKG with a dim screen and charging issue. See our technician's diagnostic and ESI process.
Mortara Eli 250 Battery Replacement Repair Case Study
Mortara Eli 250 EKG
A PBES technician diagnoses and resolves a battery issue on a Mortara Eli 250 EKG. See the steps taken in this real repair case study from Florida.
Welch Allyn 767 Wall Transformer Cord Repair Case Study
WELCH ALLYN, INC. 767 Wall Transformer, Oto/Ophthalmoscope
A PBES technician diagnoses exposed wires on a Welch Allyn 767 wall transformer and replaces the frayed cord, restoring the unit to safe operation.
Philips HeartStart AED Repair Case Study: Expired Pads
PHILLIPS MEDICAL SYSTEMS, INC. HEARTSTART AUTOMATIC EXTERNAL DEFIBRILLATOR
A PBES technician responds to a Florida hospital to replace expired defibrillator pads on a Philips HeartStart AED, ensuring the device is ready for use.
Datascope Passport 2 Repair: Settings Not Retained Case Study
DATASCOPE, INC. PASSPORT 2 Patient Monitor
A PBES repair case study on a Datascope Passport 2 patient monitor that failed to retain ECG settings and date/time after power cycles. See the diagnosis.
Welch Allyn CP150 ECG V3-V6 Signal Loss Repair Case Study
Welch Allyn CP150 ECG / EKG Machine
A real PBES repair case study. A Welch Allyn CP150 ECG lost signals on leads V3-V6. Read our technician's diagnostic process for resolving signal loss issues.
Welch Allyn 440 Stiff Power Button Found During PM | Case Study
Welch Allyn 440 Vital Signs Monitor
A PBES PM case study. A Welch Allyn 440 vital signs monitor was found to have a stiff power button during a routine preventive maintenance check.
Welch Allyn 5200 SpO2 & Battery Repair Case Study
Welch Allyn 5200 Patient Monitor
A PBES technician diagnoses SpO2 and battery failures on a Welch Allyn 5200 patient monitor. See the diagnostic steps that led to recommending a new cable and battery.
Vectra Cor Universal Smart EKG Operational Verification Case Study
Vectra Cor Universal Smart EKG EKG
A PBES technician was dispatched to a clinic to verify the function of a Vectra Cor Universal Smart EKG after a concerning automated interpretation was reported.
Welch Allyn ELI 280 No Power Repair Case Study
Welch Allyn ELI 280 ECG/EKG Machine
A PBES technician diagnoses and resolves a 'no power' issue on a Welch Allyn ELI 280 ECG machine. See the straightforward fix in this real repair case study.
Welch Allyn VSM 6000 BP Issue Repair Case Study
Welch Allyn VSM 6000 Vital Signs Monitor
A Welch Allyn VSM 6000 monitor was reported with a BP issue. See how a PBES tech performed full calibration to resolve and verify performance.
Welch Allyn CP150 Not Registering Leads Repair Case Study
Welch Allyn CP150 ECG/EKG Machine
A PBES technician diagnoses and resolves an issue with a Welch Allyn CP150 ECG machine that was not registering leads due to a loose connection. Real repair data.
Welch Allyn 767 Wall Transformer Scope Handle Repair Case Study
WELCH ALLYN, INC. 767 Diagnostic Wall Transformer
A PBES technician diagnoses and attempts to repair a Welch Allyn 767 wall transformer with exposed wires on the scope handle. A real repair case study.
Zoll R Series Defibrillator Battery Repair Case Study
Zoll R SERIES Defibrillator
A real PBES repair case study of a Zoll R Series defibrillator showing an orange battery status light. Learn how the battery was replaced and tested.
Burdick ELI280 Battery Wiring Repair Case Study
Burdick ELI280 ECG/EKG Machine
A PBES technician diagnosed and repaired a Burdick ELI280 EKG machine with disconnected battery intake wiring, restoring its power and charging functions.
Welch Allyn CP150 EKG Lead Detection Repair Case Study
Welch Allyn CP150 EKG
A verified PBES repair case study on a Welch Allyn CP150 EKG with intermittent lead detection, artifact, and inaccurate readings. See our diagnostic process.
Mortara ELI 150c Interpretation Malfunction Repair Case Study
Mortara ELI 150c ECG/EKG Machine
A PBES technician diagnoses a Mortara ELI 150c EKG reported for malfunctioning and duplicated interpretations. See how a software issue was identified.
Welch Allyn Vital Signs Monitor Repair: 'Weird Beep' Issue
Welch Allyn Vital Signs Spot Monitor
A PBES technician investigates a Welch Allyn Vital Signs spot monitor reportedly making a 'weird beep'. See the diagnostic and testing process in this case study.
Midmark Patient Scale Repair: Inaccurate Readings & Load Cell Failure
Midmark Not documented Patient Scale
Read a real PBES repair case study on a Midmark patient scale in a Florida clinic that was reading underweight and showing erratic numbers, leading to a diagnosis of load cell failure.
Welch Allyn Connex Spot NIBP Issue: Repair Case Study
Welch Allyn CONNEX SPOT MONITOR Vital Signs Monitor
A PBES technician diagnoses a blood pressure issue on a Welch Allyn Connex Spot Monitor at an urgent care. See the diagnostic steps from the service report.
McKesson ELI 280 'No Leads' Error Repair Case Study
MCKESSON ELI280 ECG/EKG Machine
A PBES technician diagnoses a McKesson ELI 280 reporting a 'no leads' error. Discover how swapped lead placement and a quick calibration resolved the issue.